Instructions specific to hazardous areainstallations(reference European ATEX Directive94/9/EC, Annex II, 1.0.6.)The following instructions apply to eq
Operation Press Display Shows Comments3. Then REC icon indicates that theinstrument is now storing dosedata in the first available datastore.4. Inspec
Operation Press Display Shows Comments1. Enable DataRecallWith the instrument stopped,pressing displays theStart Date & Time for the mostrecently
Operation Press Display Shows Comments2. Enable DataRecallWith the instrument stopped,display the Start Date & Timefor the most recently storedrun
Operation Press Display Shows Comments3. ReviewSLMMeasurementsseveraltimesThe display steps through alldetails and parameters (seeSection 9.4), starti
8.2 Set Time & DateOperation Press Display Shows Comments1. Select TimeOptionFollow on from Step 3 inSection 8.1.2. Enable TimeChangeCursors are e
Operation Press Display Shows CommentsPress to enable,use and to select thefollowing durations:Off, 5, 10, 15, 30 minutes,1, 2, 4, 8, 10, 12, 24 hours
9. Technical Information9.1 SpecificationDosimeter/Sound Exposure Meter Standards:IEC 61252: BS EN 61252, ANSI S1.25 - 1992 for dosimeters and soundex
There is no degradation in performance or loss of functionality followingthe application of electrostatic discharges.The approved configuration for re
Overload Indication:140.1 dB RMS (i.e. 0.1 dB above the top of the selected range).Maximum Peak Value:143.0 dB Peak Linear (i.e. 3.0 dB above the top
Setup OSHA MSHA DOD ACGIH IS085 IS090 METER SLMThreshold 2 90 90 x x x x x x(dB)Criterion (dB) 90 90 85 85 85 90 0 xLN%A L10 L10 L10 L10 L10 L10 L10 x
Page 2 - CEL-320/360 Getting Started
the total Avg of a person’s workplace noiseexposure averaged over a user selected(variable) period of hours and minutes.This unit is specified in the
The warranty period runs for 12 months from the date of receiptof goods, with exceptions on certain specialised components supplied byother manufactur
11. Intrinsically Safe ModelsModels in the CEL-320/360 series are available with Intrinsic Safetyapproval for use in hazardous areas.The Intrinsically
CEL-320/360 Getting Started - Page 23
Page 24 - CEL-320/360 Getting Started
CEL-320/360 SERIESUSERS GUIDEIf you want to get on and use the instruments without any backgroundinformation, skip Sections 3 - 5 and go straight to S
3. Contents4. Introduction to the DosimetersThe CEL-320 and CEL-360 Noise Dosimeters have been developed fromthe successful CEL-420/460 range of instr
high resistance to damage. Data integrity is further protected by a dosemicrophone lead which is designed to resist knocks and abrasionswithout affect
However, all soundlevel measurementparameters can be set fromthe keypad. There are threeoverlapping measurementranges: 30 - 100, 50-120 &70-140 dB
Once data saved by a CEL-360 instrument has been downloadedto a PC, the user can specify a period over which to re-calculate theprojected % dose value
Fit the CEL-4726 Couplersupplied with the calibrator (CEL-110/1 or110/2 ) into the calibrator cavity, makingsure it is firmly in contact with theshoul
6. Operation as a Dosimeter6.1 Select Dose Setup for UseOperation Press Display Shows Comments1. ChangeOperatingModere-peatedlyTo select SET UP.The fi