Casella-cel CEL-281 Manual do Utilizador Página 132

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 147
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 131
CEL-6225 Dust Cover has no measurable effect on acoustic
performance of microphone.
Electrical substitution of microphone input for
calibration:
Not recommended to be performed by unqualified personnel.
The procedure requires the microphone assembly to be replaced
by the CEL-6100 Dummy Microphone and recommends use of
an oscillator with 600 impedance.
Consult CEL for precise details.
Operating environment:
Temperature: -10 to +50 C,
(Effect on
accuracy: ±0.5 dB re 20
C at 65% RH),
Atmospheric
Pressure: 101.3 kPa ±10%
(1013 mbar, 760 mmHg),
Humidity: 30 to 90% RH (non-condensing),
(Effect on
accuracy: ±0.5 dB re 65% RH at 40
C),
Magnetic Fields: Negligible effect,
Vibration: Negligible effect.
Storage environment:
(without batteries)
Temperature: -10 to +50
C,
Humidity: 30 to 95% RH (non-condensing).
Consult the battery Manufacturer for storage conditions with
batteries.
Technical Information